Compatible in every working environment, used across pharmaceutical and biotech settings.

Benefits

  • Customised size RLAF supplied to requirement
  • 3-stage filtration for consistent air quality
  • Installation service available

Typical specification

MOC
Stainless steel 304 grade / Powder coated MS
Filtration
3-stage — HEPA (0.3 micron), pre-filter (10 micron), fine filter (5 micron)
Air Flow Direction
Recirculating laminar air flow
Air Cleanliness
Class 100, ISO Class 5
Air Velocity
90±20 cfm (average)
Noise Level
68±3 dB
Control Panel
MCB 32 AMP
Power Supply
Single phase 220V, 50Hz

Pharmaceutical applications

  • Pharmaceutical
  • Biotech

Our expertise

  • Heavy-duty blower motor assembly
  • Anti-static PVC curtain
  • PAO port for HEPA integrity testing
  • Full switch panel: mains, light, UV, blower, air/gas cock, magnehelic gauge, hour meter

Custom capacities, automation levels, and compliance packages are available. Share your process requirement and we will quote against it.